MJ-560B3 Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility. The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process are what demonstrate the inner quality of Aokai products: mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Machine is both outstanding in performance and versatile in functions.
n The Ventilator adopts microcomputer control with high-definition TFT color LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
n The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
n High-definition four-tube flowmeter, designed with stoppers for oxygen and nitrous oxide to control the output oxygen concentration.
n Reliable pneumatic transmission system.
n Economic and proper configuration of technologies, optional as required by the Customer.
n Modularized design and upgrading. Upgrading is available as required by the Customer.
Part of Anesthesia Machine
■Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
■Scope of application adult, child
■Gas source O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
■Flow meter O2: 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
■O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
■Flow rate of rapid oxygen supply 25~75L/min
■Low oxygen pressure alarm There will be sound alarm when the oxygen pressure < 0.2MPa
■Vaporizer it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
■Respiratory circuit working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
■ Respiratory Bellows bellows for adults, bellows for children
tidal volume range:0~1,500 ml
Part of Ventilator
■Display mode High-definition TFT color LCD screen display
■Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL,
SIGH, (IMV, VCV, PCV)
■ Ventilation parameters Tidal volume 50~1500 ml (20~1500 ml)
Rate 1~99 bpm
SIMV rate 2~20 bpm
I:E 2:1~1:8 (4:1~1:8)
Inspiratory trigger pressure -1.0~2.0Kpa
PEEP 0~2.0Kpa
Pressure Range 1.0~6.0Kpa
SIGH 1.5 times the inspiratory time (70-150
adjustable)
Inspiratory platform 0~50%
■ Parameters for ventilation monitoring
tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow
-volume oop,pressure-volume loop)
■Monitoring of oxygen concentration 21%~100%
■Security Alarm System
Oxygen Concentration alarm upper limit setting range 21%~100%
low limit setting range 10%~80%
Airway pressure alarm upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there
is no tidal volume input for 15 seconds
Power alarm
■ Power AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
■(Mechanical Arms)
■(End-expiratory CO2 monitoring)
*Items within brackets are optional.