Part of Anesthesia
■Main body                  High-strength engineering plastic rack, light,
beautiful and corrosion resistant
■Scope of application           adult, child
■Gas source                  O2, N2O: 0.27~0.55MPa
■Flow meter                  O2: 0.1~1.0L/min    1.1~10L/min
■Flow rate of rapid oxygen supply  25~75L/min
■Vaporizer           Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer.
■Respiratory circuit             working mode: all-close, semi-close, semi-open
                                             releasing pressure:1~4 kPa
■Respiratory bellows            bellows for adults, bellows for children 
                                               tidal volume range:0~1500 ml
 
Part of Ventilator
■Display mode                 High-definition 4.8” LCD screen display
■Ventilation  mode             IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
■Ventilation parameters          Tidal volume         50~1500ml
                                                 Rate                4~80 bpm
                                                  I:E                2:1~1:4
                                                  Pressure range        1.0~6.0 kPa
■Parameters for ventilation monitoring   tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform
■Security Alarm System
Airway pressure alarm          upper limit setting range  1.0~6.0 kPa
                                             low limit setting range    0.4~2.0 kPa
TV alarm                            upper limit setting range   0.5~2.0L
                                         low limit setting range     0.02~1.0L 
Sustained high-pressure alarm      it will give alarm when stress have consistently
                                                       been higher than 2.5 kPa
Suffocation alarm              it will give sound and light alarm if there is 
                                          no tidal volume input in 15s
Power alarm
■Power                           AC 220V±10%     50Hz±1Hz