AV-2000B2 is a multi-functional Ventilator, it is applicable for children and adults in sickroom, ICU or of emergency.
 
Ventilator
■Display mode                     High-definition 8.4” TFT color LCD screen display
■Ventilation  mode               IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
■Ventilation parameters           Tidal volume        50~1500 ml 
                                                   Rate               2~99 bpm
                                                   SIMV rate          2~20 bpm
                                                   I:E                2:1~1:8 
                                                   Inspiratory trigger pressure -1.0~2.0kPa
                                                   PEEP              0~2.0 kPa
                                                   Pressure range       1.0~6.0 kPa
                                                   SIGH     1.5 times the inspiratory time
                                                   Inspiratory platform   0~50%
■Parameters for ventilation monitoring    
       tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform 
■Monitoring of oxygen concentration    21%~100%
■Security Alarm System
Oxygen Concentration alarm      upper limit setting range 50%~100%
                                                   low limit setting range   15%~50%
Airway pressure alarm                upper limit setting range   1.0~6.0 kPa
                                                    low limit setting range    0.4~2.0 kPa
Per-minute ventilation volume alarm        upper limit setting range 3.0~30L/min
                                                                 low limit setting range  1.0~10L/min
Sustained high-pressure alarm       it will give alarm when stress have consistently 
                                                        been higher than 2.5 kPa 
Suffocation alarm              it will give sound and light alarm if there is no tidal      volume input for 15 seconds
Power alarm
■ Power                          AC 220V±10%     50Hz±1Hz
                                        UPS, Storage battery
■Mechanical Arms