Specifications
Light and audio alarm for set dose rate, symble display at the same time. continuous audio and light flashing if dose rate is more than standard.
Main specifications
1. Detector- GM tube
2. Display unit- dose rate (µSv/h or mSv/h), cumulate dose (µSv or mSV), peak value, Max. dose rate, dose exposuring time etc.
3. Dose rate: Hp(10)0.0μSv-9999mSv
cumulate dose: Hp(10)0.1Sv/h-500mSv/h
4. Error: 10%(2 mSv/h)
5. Dose rate response: ≤±20%
6,Energy range: ≤±30% 50KeV—60Co
7. Testing time: 300 second—1second convert automaticly
8. Memory: 64K for one week memory datas
9. Alarm function: Light and audio alarm for set dose rate, symble display at the same time. continuous audio and light flashing if dose rate is more than 1000 mR/h and or GM tube is blocked, sound rate is 80dB at the distance of 30cm.
10 Low battery indication
11. Power: AAA 1.5V battery 1pc, continuously working of 720 hours.
12. Working temperature: ≤±10% ,10°C~ +50°C
13. Working humidity: ≤±10% 0~ 95%RH(+35°C)
14. Size: 55(W)×75(L)×20(H)
15. Weight: 98g (battery included)